15 Aralık 2019 Pazar

HEPA / ULPA Filter İntegrity Test


# Before testing, make sure that the air speed and volume of the filters must be appropriate.
# To access the filters, existing air diffusers and airflow laminators (whatever can be applied) must be removed.
 # HEPA / ULPA filter is a photometer with an aerosol concentration of 20-80 mg / L and an aerosol concentration value indicator (e.g. HEPA).ATI TDA, 2H,2G)
# set the Aerosol photometers to 100% and zero and save the values to the table above the item
# change to the test position (downstream) and scan the total surface of the filter and the edges that keep the probe approximately 2.5 cm below the surface of the filter with a scan speed of no more than 0, 093m2/min (scan test).
* # If maximum applicable after 5 HEPA/ULPA filter tests during testing, aerosol concentration, 100% and zero control must be verified and recorded.